Jostent GraftMaster Coronary Stent Graft System (International)

GraftMaster Coronary Stent Graft comes pre-mounted on a delivery system that consists of a low profile, trackable and durable semi-compliant balloon catheter.

Features & Benefits

Technical Information

Ordering Information

Features & Benefits

 

Top


What's Your Next Move?

GraftMaster is indicated for use in the treatment of free perforations - defined as free contrast extravasation into the pericardium, in native coronary vessels or saphenous vein bypass grafts > 2.75 mm in diameter.

Top


Technical Information

Stent Specification
Stent Material: Stainless Steel 316L
Graft Material: PTFE
Wall Thickness: 0.3 mm
Expansion Range of Stent: 3.0 – 5.0 mm (standard vessel)



Delivery System Specification
Balloon Material: (semi-compliant)
Minimum Deployment Pressure: 14 atm for Ø 3.0 - 4.0 mm
12 atm for Ø 4.5 - 5.0 mm
Rated Burst Pressure: 16 atm for Ø 3.0 - 4.0 mm
14 atm for Ø 4.5 - 5.0 mm
Crimped Profile: 1.5 mm for 3.0 - 3.5 mm
1.6 mm for 4.0 mm
1.7 mm for 4.5 - 5.0 mm
Minimum Guiding Catheter Ø: 7F min ID = 0.074"
Shaft Size: 3.0F proximal 2.6F distal
Maximum Guide Wire: 0.014"

 

Top


Ordering Information

Ordering Number Stent Length
(mm)
Diameter
(mm)
Rated Burst Pressure
(bar)
210CG0930 9 3.0 16
210CG0935 9 3.5 16
210CG0940 9 4.0 16
210CG1230 12 3.0 16
210CG1235 12 3.5 16
210CG1240 12 4.0 16
210CG1630 16 3.0 16
210CG1635 16 3.5 16
210CG1640 16 4.0 16
210CG1645 16 4.5 14
210CG1650 16 5.0 14
210CG1930 19 3.0 16
210CG1935 19 3.5 16
210CG1940 19 4.0 16
210CG1945 19 4.5 14
210CG1950 19 5.0 14
210CG2630 26 3.0 16
210CG2635 26 3.5 16
210CG2640 26 4.0 16
210CG2645 26 4.5 14
210CG2650 26 5.0 14

Top