StarClose Vascular Closure System (U.S.)
Closure Redefined
Four Clicks to Close
- Easy to learn and use
- 100% Procedure Success in CLIP Study*
- Through-the-Sheath Deployment – minimize device exchanges
- Automated Closure – nitinol clip instantly closes Arteriotomy
Extravascular Closure
- Designed not to impact lumen diameter or distal bloodflow
- Mechanical Closure – hemostasis is not clot dependent
- Nitinol Clip – nothing remains in the artery
- Circumferential – provides 360° tissue apposition
Reliable Outcomes
- Safe, effective closure
- Rapid Hemostasis – CLIP study median time to hemostasis was 16.8 seconds*
- Security – hemostasis can be challenged and confirmed on the table
* based on diagnostic arm of the CLIP (Clip closure in Percutaneous procedures) clinical study which compared the safety and effectiveness for StarClose versus manual compression in 208 patients.
Ordering Information
| Catalog # | Description | Quantity |
| 14677-01 | StarClose Clip Applier(includes)
|
10 unit boxes |
Indications and Important Safety Information
StarClose
Vascular Closure System
INDICATIONS FOR USE
The StarClose Vascular Closure System is indicated for the percutaneous closure of common femoral artery access sites while reducing times to hemostasis, ambulation, and dischargeability in patients who have undergone diagnostic endovascular catheterization procedures utilizing a 5F or 6F procedural sheath.
The StarClose Vascular Closure System is indicated for use to allow patients who have undergone diagnostic endovascular catheterization procedures to ambulate and be eligible for discharge as soon as possible after device placement.
The StarClose Vascular Closure System is indicated for the percutaneous closure of common femoral artery access sites while reducing times to hemostasis and ambulation in patients who have undergone interventional endovascular catheterization procedures utilizing a 5F or 6F procedural sheath.
CONTRAINDICATIONS
The StarClose Vascular Closure System is contraindicated for use in patients with known hypersensitivity to nickel-titanium.
WARNINGS
Do not use the StarClose Vascular Closure System if the packaging or sterile barrier has been previously opened or damaged or if the components appear to be damaged or defective.
DO NOT RESTERILIZE OR REUSE. The StarClose Vascular Closure System and accessories are intended for single use only.
Do not use the StarClose Vascular Closure System if the sterile field has been broken where bacterial contamination of the sheath or surrounding tissues may have occurred, since such a broken sterile field may result in infection.
Do not use the StarClose Vascular Closure System if the puncture site is located above the most inferior border of the inferior epigastric artery (IEA) and/or above the inguinal ligament based upon bony landmarks, since such a puncture site may result in a retroperitoneal hematoma. Perform a femoral angiogram to verify the location of the puncture site.
Do not use the StarClose Vascular Closure System if the puncture is through the posterior wall or if there are multiple punctures, since such punctures may result in a retroperitoneal hematoma.
Do not use the StarClose Vascular Closure System if the puncture site is located in the superficial femoral artery or the profunda femoris artery, since such puncture sites may result in a pseudoaneurysm, intimal dissection, or an acute vessel closure (thrombosis of small artery lumen). Perform a femoral angiogram to verify the location of the puncture site.
PRECAUTIONS
1. The StarClose Vascular Closure System should be used only by operators trained in diagnostic and interventional catheterization procedures who have been certified by an authorized representative of Abbott Vascular Inc.
2. The StarClose Vascular
Closure System is provided sterile and
non-pyrogenic in unopened undamaged packaging. Products are sterilized with
ethylene oxide and intended for single use only. Do not resterilize. Store in a
cool, dry place.
3. Prior to use, inspect the StarClose Vascular Closure System to ensure that the sterile packaging has not been damaged during shipment. Examine all components prior to use to verify proper function. Exercise care during device handling to reduce the possibility of accidental device breakage.
4. As with catheter based procedures, infection is a possibility. Observe sterile technique at all times when using the StarClose Vascular Closure System. Employ appropriate groin management post procedure and post hospital discharge to prevent infection.
5. Use a single wall puncture technique. Do not puncture the posterior wall of the artery.
6. Do not use the StarClose Vascular Closure System to close vessels with diameters less than 5 mm.
7. Do not deploy the Clip in areas of calcified plaque.
8. The StarClose Vascular Closure System can be used ONLY with the StarClose Exchange System (included in the StarClose Vascular Closure System packaging).
9. Do not advance the StarClose Vascular Closure device against resistance until the cause of that resistance has been determined. Excessive force used to advance or torque the StarClose device should be avoided, as this may lead to significant vessel damage and/or breakage of the device, which may necessitate interventional and/or surgical removal of the device and vessel repair. Refer to Instructions for Use for additional Precaution information.
ADVERSE EVENTS
Potential adverse events that could be associated with use of this device include:
• Major Vascular Complications
– Vascular Injury Requiring Repair
– Surgery
– Angioplasty
– Ultrasound Guided Compression
– Thrombin injection or Other Percutaneous Procedure
– New Ipsilateral Lower Extremity Ischemia
• Access Site-related Bleeding Requiring Transfusion
• Access Site-related Infection Requiring Intravenous Antibiotics or Prolonged Hospitalization
• Access Site-related Nerve Injury Requiring Intervention
• Death
• Minor Vascular Complications
– Pseudoaneurysm
– Arteriovenous Fistula
– Hematoma (≥ 6cm)
– Late access site-related bleeding
– Transient lower extremity ischemia
– Ipsilateral deep vein thrombosis
– Transient access site-related nerve injury
– Access site-related vessel injury
– Access site wound dehiscence
– Ultrasound guided Thrombin Injection
– Re-bleeding at time of
first ambulation, requiring ≥ 30 min. for
re-hemostasis
– Access site-related bleeding requiring ≥ 30 minutes to re-achieve hemostasis
– Localized access site infection treated with IM or oral antibiotics
REPUNCTURE THROUGH STARCLOSE CLIP
The safety of repuncture at any time through any part of a previously deployed StarClose Clip, and the safety of subsequent closure of this repuncture using the StarClose Vascular Closure System, have not been fully established. The following information is provided to assist the operator in assessing the possible risks that may be associated with such repuncture and repuncture closure, which include Clip dislodgement, Clip embolization, and bleeding. Studies were performed in a simulated setting because a clinical trial would not be adequate in testing the worst case scenario since the likelihood of hitting the Clip in the clinical setting is very low. Please consult IFU for complete information regarding the studies performed.
AP2929609 Rev. A