This web site is not optimized for Internet Explorer 6. You may notice
decreased functionality compared to other web browsers. Please follow the
link below to upgrade to a new version of Internet Explorer.

The "Yes" link below will take you out of the Abbott Laboratories family of websites.
Links which take you out of Abbott Laboratories worldwide web sites are not under the
control of Abbott Laboratories, and Abbott Laboratories is not responsible for the
contents of any such site or any further links from such site. Abbott Laboratories is
providing these links to you only as a convenience, and the inclusion of any link does
not imply endorsement of the linked site by Abbott Laboratories.

Do you wish to leave this site?
 
Coronary Dilatation Catheters

JOCATH MERCURY PTCA Catheter

Indications & Important Safety Information ››
 
Overview

Abbott Vascular Devices brings you a PTCA balloon catheter for even the most complex procedures.

The unique shaft construction of Mercury ensures full control of the catheter from hub to tip - proximally providing the push where performance is needed, while distally displaying the flexibility and low profile for excellent lesion crossability.*

Mercury has a versatile design making it suitable for a wide range of applications. Its unique controlled compliance characteristics will make it your first choice every time.

mercury

* Tests performed by and data on file at Abbott Vascular.

Specifications
Balloon Material:
bm2
Balloon Compliance:
Semi-compliant
Shaft Diameter: 2.0F proximal 2.7F distal
Shaft Length: 140-145 cm
Lesion Entry Profile:
0.017"
Nominal Pressure:
8 bar
Rated Burst Pressure:
16 bar
Average Burst Pressure:
22 bar
Coating: HYDREX Coating System
Min. Guiding Catheter:
5F (0.058")

 

 

 

JOCATH MERCURY Compliance Charts*

Pressure (bar)Balloon Diameter (mm)
1.502.002.25
2.50
2.75
3.00
3.25
3.50
4.00
4 1.43 1.91 2.15 2.39 2.63 2.87 3.11 3.34 3.82
5 1.45 1.93 2.18 2.42 2.66 2.90 3.14 3.38 3.86
6 1.47 1.96 2.20 2.45 2.69 2.93 3.18 3.42 3.91
7 1.48 1.98 2.23 2.47 2.72 2.97 3.22 3.46 3.95
8 1.50 2.00 2.25 2.50 2.75 3.00 3.25 3.50 4.00
9 1.52 2.03 2.28 2.53 2.78 3.04 3.29 3.54 4.04
10 1.53 2.05 2.31 2.56 2.81 3.07 3.33 3.58 4.09
11 1.55 2.07 2.33 2.59 2.84 3.10 3.36 3.62 4.13
12 1.57 2.10 2.36 2.61 2.87 3.14 3.40 3.66 4.18
13 1.59 2.12 2.38 2.64 2.91 3.17 3.44 3.70 4.22
14 1.60 2.14 2.41 2.67 2.94 3.21 3.48 3.74 4.27
15 1.62 2.16 2.44 2.70 2.97 3.24 3.51 3.78 4.31
16 1.64 2.19 2.46 2.73 3.00 3.27 3.55 3.82 4.36
17 1.65 2.21 2.49 2.75 3.03 3.31 3.59 3.86 4.40
18 1.67 2.23 2.51 2.78 3.06 3.34 3.62 3.90 4.45
19 1.69 2.26 2.54 2.81 3.09 3.38 3.66 3.94 4.49
20 1.70 2.28 2.57 2.84 3.12 3.41 3.70 3.98 4.54

† Nominal pressure
‡ Rated burst pressure
* Tests performed by and data on file at Abbott Vascular.

Ordering Information
Order NumberBalloon LengthBalloon DiameterNo. of Radiopaque Markers
M31115 11 mm 1.50 mm 1
M31120 11 mm 2.00 mm 2
M31125 11 mm 2.50 mm 2
M31130 11 mm 3.00 mm 2
M31135 11 mm 3.50 mm 2
M31140 11 mm 4.00 mm 2
M31415 14 mm 1.50 mm 1
M31420 14 mm 2.00 mm 2
M31425 14 mm 2.50 mm 2
M31430 14 mm 3.00 mm 2
M31435 14 mm 3.50 mm 2
M31440 14 mm 4.00 mm 2
M32015 20 mm 1.50 mm 1
M32020 20 mm 2.00 mm 2
M32022 20 mm 2.25 mm 2
M32025 20 mm 2.50 mm 2
M32027 20 mm 2.75 mm 2
M32030 20 mm 3.00 mm 2
M32032 20 mm 3.25 mm 2
M32035 20 mm 3.50 mm 2
M32040 20 mm 4.00 mm 2

AP2932857 Rev. A

 