Coronary Stent Grafts

JOSTENT GRAFTMASTER Coronary Stent Graft System

Indications & Important Safety Information ››
 
Overview

graftmasterGRAFTMASTER Coronary Stent Graft comes pre-mounted on a delivery system that consists of a low profile, trackable and durable semi-compliant balloon catheter.

 

 

Tests performed by and data on file at Abbott Vascular.
Photos taken by and on file at Abbott Vascular.

Features
graftmaster 2

Stent Delivery System

  • Easy, safe and accurate delivery of the GraftMaster Coronary Stent Graft
  • Uniform expansion of the stent graft due to unique four-fold wrap of balloon
  • Two markers for accurate graft placement
  • A tapered zero-clearance tip design for crossability and trackability
  • Lesion entry profile of 0.018"
  • HYDREX Coating System on the tip and on the distal shaft
sandwichedfold

Stent Graft

  • Constructed using a sandwich technique with an ultra thin layer of expandable PTFE is placed between two stainless steel Jostent stents
  • Low crimped profile which is comparable to conventional stent systems
  • PTFE graft material is constructed to allow for expansion in one direction, while remaining rigid in the other and when fully expanded Jostent Coronary Stent Graft retains its radial strength to resist vessel wall or lesion pressure
What's Your Next Move?
fl 2 fl 3
GraftMaster is indicated for use in the treatment of free perforations - defined as free contrast extravasation into the pericardium, in native coronary vessels or saphenous vein bypass grafts > 2.75 mm in diameter.

 

Photos taken by and on file at Abbott Vascular.
Tests performed by and data on file at Abbott Vascular.

Specifications
Stent Specification
Stent Material:
Stainless Steel 316L
Graft Material:
PTFE
Wall Thickness:
0.3 mm
Expansion Range of Stent:
3.0 - 5.0 mm
Delivery System Specifications
Balloon Material:
(semi-compliant)
Minimum Deployment Pressure:
14 atm for Ø 3.0 - 4.0 mm
12 atm for Ø 4.5 - 5.0 mm
Rated Burst Pressure:
16 atm for Ø 3.0 - 4.0 mm
14 atm for Ø 4.5 - 5.0 mm
Crimped Profile:
1.5 mm for 3.0 - 3.5 mm
1.6 mm for 4.0 mm
1.7 mm for 4.5 - 5.0 mm
Shaft Size:
2.7F proximal / 2.0F distal (for 3.0-4.0 mm) 

3.0F proximal / 2.0F distal (for 4.5-5.0 mm)
Maximum Guide Wire:
0.014"

 

Tests performed by and data on file at Abbott Vascular.

Ordering Information
Ordering NumberStent Length
(mm)
Diameter
(mm)
Rated BurstPressure
(bar)
210CG0930 9 3.0 16
210CG0935 9 3.5 16
210CG0940 9 4.0 16
210CG1230 12 3.0 16
210CG1235 12 3.5 16
210CG1240 12 4.0 16
210CG1630 16 3.0 16
210CG1635 16 3.5 16
210CG1640 16 4.0 16
210CG1645 16 4.5 14
210CG1650 16 5.0 14
210CG1930 19 3.0 16
210CG1935 19 3.5 16
210CG1940 19 4.0 16
210CG1945 19 4.5 14
210CG1950 19 5.0 14
210CG2630 26 3.0 16
210CG2635 26 3.5 16
210CG2640 26 4.0 16
210CG2645 26 4.5 14
210CG2650 26 5.0 14

 

Tests performed by and data on file at Abbott Vascular.

AP2932843 Rev. A

 