Coronary Dilatation Catheters

NC MERCURY PTCA Catheter

Indications & Important Safety Information ››
 

 

Overview

nc mercuryReliable Performance Under High Pressure

The high pressure NC MERCURY PTCA catheter leverages the high-end technology of the Mercury platform.

Features
Chart

Superb Deployment*

  • Maximized wall apposition due to non-compliant balloon behavior
  • Minimal balloon overhang designed to limit vessel injury during high pressure dilatation

Efficient Deliverability*

  • Minimal tip flaring
  • Procedural efficiency maintained through a fast deflation time
  • Efficient access due to a low entry profile

 

*Tests performed by and data on file at Abbott Vascular.

Specifications

Compliance Chart*

barkPa2.502.753.003.253.503.754.004.254.50
4 400 2.35 2.59 2.82 3.06 3.29 3.53 3.76 4.00 4.23
5 500 2.38 2.61 2.85 3.09 3.33 3.56 3.80 4.04 4.28
6 600 2.40 2.64 2.88 3.12 3.36 3.60 3.84 4.08 4.32
7 700 2.43 2.67 2.91 2.15 3.40 3.64 3.88 4.12 4.37
8 800 2.45 2.70 2.94 3.19 3.43 3.68 3.92 4.17 4.41
9 900 2.48 2.72 2.97 3.22 3.47 3.71 3.96 4.21 4.46
10 1000 2.50 2.75 3.00 3.25 3.50 3.75 4.00 4.25 4.50
11 1100 2.51 2.76 3.02 3.27 3.52 3.77 4.02 4.27 4.52
12 1200 2.53 2.78 3.03 3.28 3.54 3.79 4.04 4.30 4.55
13 1300 2.54 2.79 3.05 3.30 3.56 3.81 4.06 4.32 4.57
14 1400 2.55 2.81 3.06 3.32 3.57 3.83 4.08 4.34 4.60
15 1500 2.57 2.82 3.08 3.34 3.59 3.85 4.11 4.36 4.62
16 1600 2.58 2.84 3.10 3.35 3.61 3.87 4.13 4.39 4.64
17 1700 2.59 2.85 3.11 3.37 3.63 3.89 4.15 4.41 4.67
18 1800 2.61 2.87 3.13 3.39 3.65 3.91 4.17 4.43 4.69
19 1900 2.62 2.88 3.14 3.41 3.67 3.93 4.19 4.45 4.71
20 2000 2.63 2.90 3.16 3.42 3.69 3.95 4.21 4.48 4.74
21 2100 2.65 2.91 3.18 3.44 3.71 3.97 4.24 4.50 4.77
22 2200 2.66 2.93 3.20 3.46 3.73 4.00 4.26 4.53 4.80
23 2300 2.68 2.95 3.22 3.48 3.75 4.02 4.29 4.56 4.82

 

 

Technical Information*

Balloon Material: OptiLiance, non-compliant
Balloon Markers: Platinum/Iridium, embedded
Shaft Diameter: 2.0F proximal 2.7F distal
Shaft length: 140 - 145cm
Lesion Entry Profile: 0.017"
Nominal Pressure: 10 bar
Rated Burst Pressure: 20 bar
Coating: HYDREX Coating System
Min. Guiding Catheter: 5F (0.058") for 2.50-4.25mm 6F (0.068") for 4.5mm

 

† Nominal pressure
‡ Rated burst pressure
* Tests performed by and data on file at Abbott Vascular.

Ordering Information
Order NumberStent Length (mm)Nominal Stent Diameter(mm)
82123-01 10 2.50
82124-01 10 2.75
82125-01 10 3.00
82126-01 10 3.25
82127-01 10 3.50
82128-01 10 3.75
82129-01 10 4.00
82130-01 10 4.25
82131-01 10 4.50
82123-02 15 2.50
82124-02 15 2.75
82125-02 15 3.00
82126-02 15 3.25
82127-02 15 3.50
82128-02 15 3.75
82128-02 15 4.00
82130-02 15 4.25
82131-02 15 4.50
82123-03 20 2.50
82124-03 20 2.75
82125-03 20 3.00
82126-03 20 3.25
82127-03 20 3.50
82128-03 20 3.75
82128-03 20 4.00
82130-03 20 4.25
82131-03 20 4.50

AP2932858 Rev. A

 

Indications and Important Safety Information

Top